Specifications
| Display mode | High-definition 5.7” LCD screen display |
| Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH |
| Ventilation parameters | Tidal volume 50~1500ml |
| Rate 2~99bpm | |
| SIMV rate 2~20bpm | |
| I:E 2:1~1:8 | |
| Inspiratory trigger pressure -1.0~2.0 kPa | |
| PEEP 0~2.0 kPa | |
| Pressure range 1.0~6.0 kPa | |
| SIGH 1.5 times the inspiratory time | |
| Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate,
SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure |
| Security Alarm System (Oxygen Alarm) | |
| Airway pressure alarm | upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa |
| Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min |
| Sustained high-pressure alarm | it will give alarm when stress have consistently
been higher than 2.5 kPa |
| Suffocation alarm | it will give sound and light alarm if there is no
tidal volume input for 15 seconds |
| Power alarm | |
| Power | AC 220V±10% 50Hz±1Hz
UPS, Storage battery |
| Mechanical Arms | |







Reviews